Benefits and Features
- Quality construction to ensure product reliability
- Rugged stainless-steel case protecting the piezoresistive sensor
- Compatible with most Campbell Scientific data loggers
- Fully temperature-compensated
- Low-power sleep state between measurements to reduce power consumption
- Optional weighted nose cone to facilitate submersion
- Optional NPT nose cone to enable usage in closed-pipe applications
- Quick shipment after receipt of order (ARO)
Detailed Description
The CS451 consists of a piezoresistive sensor and a temperature sensor housed in a 316L stainless-steel case. It has a rugged Hytrel cable that remains flexible, even under harsh environmental conditions. The cable incorporates a vent tube to compensate for atmospheric pressure fluctuations. The vent tube terminates inside a desiccant tube, which prevents water vapor from entering the inner cavity of the transducer. The sensor ships with a desiccant tube that can be replaced in the field.
The CS451 has several pressure range options and two accuracy options (see Ordering Info). The standard accuracy option provides ±0.1% FS TEB over the 0° to 60°C temperature range. The high accuracy option provides ±0.05% FS TEB over the 0° to 60°C temperature range and includes a calibration certificate. TEB is the combined errors due to nonlinearity, hysteresis, non-repeatability, and thermal effects over the compensated temperature range, per ISA S51.1. Please note that the high accuracy option is not available for the 0 to 2.9 psig range option.
Campbell Scientific offers the A150 Desiccated Case that allows the CS451 to be connected to a prewired enclosure (see Ordering Information).
Specifications
Measurement Time | < 1.5 s |
Output Options | SDI-12 (version 1.3) 1200 bps; RS-232 9600 bps |
Water-Level Resolution | 0.0035% FS |
Worst-Case Temperature Resolution | 0.006°C |
Dry Storage Temperature Range | -40° to +100°C WARNING: Sensor could be damaged if encased in frozen ice. |
Operating Temperature Range | 0° to 60°C WARNING: Sensor could be damaged if encased in frozen ice. |
Temperature Accuracy | ±0.2°C |
Overpressure | 2 x pressure range |
Power Requirements | 6 to 18 Vdc |
Cable Type | Hytrel Jacket, five conductor, 26 AWG |
NPT Fitting | 1/4-in. NPS |
Top Cone Material | Delrin |
Body Material | 316L stainless steel |
Element Material | 316L stainless steel |
Distance |
|
Ingress Protection | Exceeds IP68 |
Diameter | 21.34 mm (0.84 in.) |
Cable Outer Diameter |
|
Length | 213.36 mm (6.875 in.) |
Cable Weight | 0.0421 kg/m (0.0283 lb/ft) |
Weight | 0.17 kg (0.37 lb) without cable |
Air Gap |
|
Standard & Weighted Nose Cone | 0.653 cm (0.257 in.) |
NPT Fitting | 2.72 cm (1.07 in.) |
Power Consumption |
|
Quiescent | < 50 µA |
Measurement/Communication | 8 mA (1 s measurement) |
Maximum | 40 mA |
Measurement Ranges at Fresh Water Depths |
|
0 to 2.0 m (6.7 ft) |
|
0 to 5.1 m (16.7 ft) |
|
0 to 10.2 m (33.4 ft) |
|
0 to 20.4 m (67 ft) |
|
0 to 50.9 m (167 ft) |
|
0 to 102 m (334.5 ft) |
|
Accuracy |
|
Standard Accuracy Option | ±0.1% full-scale-range TEB Total Error Band (TEB) includes the combined errors due to nonlinearity, hysteresis, nonrepeatability, and thermal effects over the compensated temperature range, per ISA S51.1. |
High Accuracy Option | ±0.05% full-scale-range TEB The high accuracy (±0.05% full-scale range) option is not available for the 0 to 2.9 psig range option.Total Error Band (TEB) includes the combined errors due to nonlinearity, hysteresis, nonrepeatability, and thermal effects over the compensated temperature range, per ISA S51.1. |
Maximum Cable Length |
|
SDI-12 |
|
RS-232 | 60 m (200 ft) |