The SIF-111-SS is a fast response model infrared radiometer with a 0.2 second response time. The analog sensor is un-amplified and has a voltage output and high-quality cable terminating in pre-tinned pigtail leads for easy connection to dataloggers and controllers. The sensor has a 22° half-angle field of view. Typical applications include plant canopy temperature measurement for use in plant water status estimation, phenotyping, phenomics, road surface temperature measurement of determination of icing conditions, and terrestrial surface (soil, vegetation, water, snow) temperature measurement in energy balance studies. Sensor includes IP68 marine-grade stainless-steel cable connector 30 cm from head to simplify sensor removal and replacement for maintenance and recalibration. The SIF-111-SS is designed to be mounted with an Apogee AM-250 commercial-grade mounting bracket.
FOR MORE PRODUCT INFORMATION, CHECK OUT THESE LINKS
• Product Manual
• Specification Sheet
• Technical Drawing
Approximate Sensitivity
15 µV per CÂ
Output from Thermopile
Approximately -0.825 to 0.825 mV for a temperature difference from -55 to 55 C
Output from Thermistor
2500 mV excitation (typical, other voltages can be used)
Input Voltage Requirement
0 to 2500 mV (typical, depends on input voltage)
Calibration Uncertainty (-40 to 80 C)
0.5 C
Measurement Repeatability
Less than 0.05 C
Stability (Long-term Drift)
Less than 2 % change in slope per year when germanium filter is maintained in a clean condition
Response Time
0.2 s, time for detector signal to reach 95 % following a step change
Field of View
22° half angle
Spectral Range
8 to 14 µm; atmospheric window
Operating Environment
-50 to 80 C; 0 to 100 % relative humidity (non-condensing)
Dimensions
23 mm diameter, 60 mm length
Mass
190 g (with 5m of lead wire)
Cable
5 m of six conductor, shielded, twisted-pair wire; TPR jacket (high water resistance, high UV stability, flexibility in cold conditions); pigtail lead wires
Warranty
4 years against defects in materials and workmanship